Modular polarimeter MCP 4100 / 5100/ 5500

brand

Technical specifications

  MCP 4100 MCP 5100 MCP 5500
Specifications
Measuring Scales °Optical Rotation, °Optical Rotation (cell-length corrected), °Specific Rotation,°Specific Rotation (cell-length corrected), % concentration (g/100 mL, g/L, g/100 cm³, kg/m³), °International Sugar Scale (not temperature-compensated), mathematic functions, and user-definable scales
Measuring range ±89.9 °OR ±89.9 °OR ±89.9 °OR
Accuracy ±0.003 °OR across entire range ±0.002 °OR across entire range <0.002 °OR across entire range
Repeatability ±0.002 °OR ±0.002 °OR ±0.001 °OR
Resolution 0.001 °OR  0.0001 °OR   0.0001 °OR  
Available wavelenghts 589 nm standard
365, 405, 436, 546, 578, 633, 880 nm opt.
589 nm standard
365, 405, 436, 546, 578, 633, 880 nm opt.
589 nm standard
365, 405, 436, 546, 578, 633, 880 nm opt.
Temp. control range Peltier temperature control
10 - 45 °C 
Peltier temperature control
10 - 45 °C 
Peltier temperature control
10 - 45 °C 
Temp. accuracy ±0.1 °C ±0.1 °C ±0.03 °C
Temp. control speed 2 °C sample temperature gradient eliminated in less than 30 seconds.
(100 mm stainless steel cell, re-filled with water)
2 °C sample temperature gradient eliminated in less than 30 seconds.
(100 mm stainless steel cell, re-filled with water)
2 °C sample temperature gradient eliminated in less than 30 seconds.
(100 mm stainless steel cell, re-filled with water)
21 CFR Part 11 compliance yes yes yes
User Assistance Features
FillingCheck™ camera  no yes yes
Automatic detection of sample cell and quartz standard by Toolmaster™ function yes
wireless Toolmaster™ technology
yes
wireless Toolmaster™ technology
yes
wireless Toolmaster™ technology
Automated calibration / adjustment process by Toolmaster™ function yes yes yes
Separate drying facility for sample cell (air pump) no optional yes
Data Management
Data managment software for multiple instruments (AP Connect) optional optional optional
Desktop software controlled optional optional optional
Engineering
Light source LED LED LED
Max. sample cell length 200 mm 200 mm 200 mm